Technical Specifications
Pump Type: Drip type infusion pump
Pumping Way: Peristalic fingers
Adjustable Drop Rate Range: 1~400 drops/min (step: 1 drop/min)
Adjustable Volume Rate Range: 1~1200 ml/h (when 1~99.9 ml/h, step: 0.1 ml/h; Greater than 99.9 ml/h, step: 1 ml/h)
Flow Precision: Within±3% (using the IV set appointed by the manufacturer or calibrated high quality IV set) Mechanical precision: Within±2%.
Purge Rate: 800 ml/h
KVO Rate: 1 ml/h (1~300 ml/h);
3 ml/h (greater than 300 ml/h);
Infusion Volume Range: 1 ml~9999.9 ml (step: 0.1 ml)
Maximum Accumulated Volume: 9999.9 ml (step: 0.1 ml)
Time Range: 1 min~9999 min (step: 1 min)
Occlusion Alarm Threshold: High: 800 mmHg±200 mmHg (106.7kPa±26.7kPa)
Medium: 500 mmHg±100 mmHg (66.7kPa±13.3kPa)
Low: 300 mmHg±100 mmHg (40.7kPa±13.3kPa)
Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation,
Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle.
Air Bubble Detector: Method: Ultrasonic wave, sensitivity: 25 μL
Fuse: F1AL/250V ("F" indicate fast, "L" indicate low breaking capacity),
2 PCS (installed inside pump).
Power Supply: AC 85~265V, 50/60Hz;
Internal Battery: 11.1V rechargeable Li-ion battery. Capacity: 2000mAh;
The pump can work more than 4 hours at the flow rate of 25 ml/h after charging for 8 hours (Medium rate specified by GB 9706.27-2005).
Power: 40VA Weight: 2.2KG
Size: 188 mm(L)198 mm(W)228 mm(H)
Environment Requirements: Transport
Ambient Temperature: -30ºC~+55ºC
Relative Humidity: 20%~95% (non-condensing)
Atmosphere Pressure: 70kPa~106kPa