Features
Technical Data
Measuring parameter
Measuring range:
Turbidity: 0.0-4.0 / 40.0 / 400.0 / 1000.0 NTU
SS: 0.0-10.0 / 100.0 / 1000.0 / 2000.0 mg/L
Resolution: 0.1 NTU, 0.1 mg/L
Accuracy: ±2 f.s
Input
Measuring sensor: innoSens 810T turbidity/ suspended solids sensor
Output
Analog Output 1:
4-20 mA isolated current output for suspended solids sensor, it can be made to represent any segment of the measuring scale
Load:Max. 600Ω
Relay 1&2 (Alarm):
Operation: ON/OFF
Selectable action : High / Low
Set Point: High/10.0 g/L; Low/4.0 g/L (adjustable)
Hysteresis: 0.01 g/L (adjustable)
Relay contacts: 5A/250VAC;5A/30VDC
Relay 3 (Wash Relay)
Wash interval: 0.1~1000hour
Wash time: 1~1000s
Relay contacts: 5A/250VAC;5A/30VDC
Interface
RS 485 Modbus RTU
Baud rate: 9600 bps
Data format: 8 bit
General Information
Power supply:90 - 260V AC, 50/60Hz; 24VDC(Optional)
Temperature:
Operating temperature: 0 - 60 ℃
Storage temperature: -20 - 70 ℃
Humidity: Max. 90% rH at 40℃ (non-condensing)
LCD display: Suspended solids value and relay status
Language: Chinese / English
Passcode: Set mode: 0022, Calibration mode: 0011
Protection grade: IP65
Dimensions: 160 166 73mm
Installation: Wall Mounting
Weight:1Kg
Sensor Information
Type: innoSens 810T
Material: 316L/PVC shell, special optical glass
Size:
Diameter: 42mm
Length: 210mm
Connection: 1”GAS
Cable:10 m
Max. Pressure: 4 bar
Protection grade: IP68