GDYJ-501 is used to measure breakdown voltage BDV of all transformer oil, dielectric oil, electrical oil, insulating oil, etc.
Introduction:
GDYJ-501 China Cheap Price IEC60156 Transformer Oil BDV Test Set is developed according to IEC156 Insulating Oil Dielectric Strength Testing Method;
LCD screen can display test procedure and test result. All the human interaction operations complete by the rotating mouse.
It is very easy for customers to measure the breakdown voltage of all kinds of oil by it.You can print the test result by the micro printer internal the tester. Of course, you can upload the test data to PC from RS232 for data analyzing. The special circuit design makes the tester has a good performance at stability and anti-noise.
Parameters
Output voltage AC 0~80KV/AC 0~100KV
Rate of voltage rise 2KV/s±10%
Measuring precision 3%RDG+0.3%FS
Test times setting 1~9
Mixing time setting 0~99 second
Waiting time setting 0-9 minute 59second
Maximum storage capacity 100 groups of test data
Test standards IEC60156
Display 256*128 LCD
Rated power supply AC 220V±10% 50Hz/60Hz
Environment temperatures 0-50°C
Relative humidity ≤80%RH
Size 420×380×410mm
Weight 28kg
Functions:
1.Breakdown test of user-defined;
2.Proof test;
3.Calibration interface for customers;
4.Unloading data to PC and making word report automatically;
Result print.